Event description: per cnf, it was reported that, the hydrophilic coating was not enough, and the using was not smooth.The patient condition following procedure was stable.Procedure outcome: completed with another of same device.What was the patient condition following procedure? stable.When was the problem noticed: unpacking.Where did the problem occur? outside the patient.Event resolved? yes.Image provided 19 november 2025.
|
|
As received, the specimen consisted of one-1 each pkg-hydro gw stf std s 150-035; returned reloaded into the dispenser assembly, and double-bagged within "zip-lock" style poly biohazard pouches.The specimen presented an overall length of approximately 148.5cm and a finished diameter of.03410" to.03315".A gage bushing certified to be.0350" passed over the length of the specimen with no more effort than the mass of the bushing sliding down the wire shaft unaided, except by gravity.Note all diameter measurements were acquired with a non-contact, toolmakers scope after allowing the specimen device to become dry after hydrating the specimen device with blood-bank saline.Upon flushing the dispenser hoop with blood-bank saline, the specimen was easily removed and subjected to visual and tactile examination.The specimen coating appeared visually and tactilely consistent when examined at 1x - 18 inches, unaided, wet.The specimen presented areas/ witness marks of skived/cut damage to the polymer jacket material, including metallic core wire exposure.The damage was located 38cm to 43cm from the distal tip.The specimen also presented large radius bend damage at 8cm and 25cm from the distal tip.Except where noted, the specimen device appeared visually and dimensionally correct.The customer-provided image shows the guidewire within its dispenser assembly, exhibiting bend damage at an indeterminate distance from the tip.However, the reported claim that, "the hydrophilic coating was not enough, and the using was not smooth," could not be substantiated based on the image.Note, the skived/cut damage was not mentioned in the complaint narrative nor captured in the customer supplied image.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing the production operators are instructed to 100% visually and tactilely inspect for any obvious defect, which includes a tactile examination of the entire length of each wire.In addition, during packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.The nature and the extent of the skived/cut damage, as well as the bend damage near the distal tip, is representative of applying excessive and/or forceful manipulation to the device while removing it from the dispenser hoop without proper hydration.As noted in the dfu operational instructions: prior to use: carefully examine the unit to verify that neither the contents nor the sterile package has been damaged in shipment.The zipwire hydrophilic guidewire is a delicate instrument and should be handled carefully.Prior to use and when possible, during the procedure, inspect the wire carefully for bends or kinks which may have occurred.To activate hydrophilic coating: - before removing the zipwire hydrophilic guidewire from the protective hoop (by its distal end), inject physiological saline solution into the hub end of the holder in order to completely wet the surface of the wire.Use of alcohol, antiseptic solutions, or other solvents must be avoided, as they may adversely affect the surface of the wire.Do not wipe the zipwire hydrophilic guidewire with dry gauze.Before inserting the zipwire hydrophilic guidewire through a catheter, fill the catheter with physiological saline solution.As indicated in the directions for use (dfu) precautions: - the surface of the zipwire hydrophilic guidewire is not lubricious unless it is wet.Before taking it out of its holder and inserting it through a catheter, fill the holder and the catheter with sterile physiological saline solution.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided and the evidence presented, it appeared that handling factors have contributed to the event as reported.Lake region medical reviewed the associated risk documentation relative to this product and confirmed that this incident is listed and that it is trending within the established occurrence threshold.Lake region's effectiveness of design verification activities brings the overall risk down to as far as possible.If any further relevant information provided, a follow up medwatch report will be submitted.Note: although annex g instructions state that with stand-alone devices annex g term should not be used, code 4755 was selected due to component code (g) showing up on the missing data report.
|