Model Number 1098834
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received a work order reporting that the simplygo device experienced an equipment source failure, and the complaint was confirmed.Additionally, found the main pca was damaged, melted filter, and damaged canister.There was no report of serious or permanent harm or injury.There was no report of medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLYGO
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23898641
Report Number2518422-2025-113288
Device Sequence Number4990005
Product Code CAW
UDI-Device Identifier00606959028838
UDI-Public(01)00606959028838
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K111885
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1098834
Device Catalogue Number1098834
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2025
Initial Report FDA Received Date12/26/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/20/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-