RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
Philips received a complaint from the customer on the v60 ventilator indicating that the liquid crystal display (lcd) was found to be defective during testing.There was no patient involvement at the time the issue was discovered.No patient or user harm was reported.A field service engineer (fse) was dispatched onsite to evaluate and repair the device.Upon arrival, the fse replaced the user interface (ui) assembly, after which the issue was resolved.The device passed the required performance verification tests per philips standard and was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key23899060
Report Number2518422-2025-059106
Device Sequence Number18739148
Product Code MNT
UDI-Device Identifier00884838009868
UDI-Public(01)00884838009868
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K102985
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberV60
Device Catalogue Number1053615
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2025
Initial Report FDA Received Date12/26/2025
Date Device Manufactured07/30/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-