Model Number H74939135302010
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and moderately calcified vessel below the knee.A 3mm x 20mm x 144cm coyote balloon catheter was advanced for dilatation.However, during first inflation at 6 atmospheres for 15 seconds, the balloon ruptured.The device was removed, and a pinhole was noted.The procedure was completed with a different device.No patient complications as a result of this event and the patient condition were good post-procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE? ES
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key23899388
Report Number2124215-2025-93239
Device Sequence Number5029955
Product Code LIT
UDI-Device Identifier08714729767237
UDI-Public(01)08714729767237(17)270503(10)33957018
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093636
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH74939135302010
Device Catalogue NumberH74939135302010
Device Lot Number0033957018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2025
Initial Report FDA Received Date12/26/2025
Date Device Manufactured05/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-