• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSION ORTHOPEDICS, INC. 2.8MM GLENOID DRILL SHAFT; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSION ORTHOPEDICS, INC. 2.8MM GLENOID DRILL SHAFT; BIT, DRILL Back to Search Results
Catalog Number DRS-0926-001
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a reverse total shoulder, using the peripheral baseplate reamer, the doctor noticed something off with the driver while spinning, pulled it out of the joint and the spring and collar inside the 2.8mm glenoid drill shaft fell out on the floor.The attachment was still in the t-handle and caught the outer black shaft.Nothing fell inside the patient; it was confirmed with x-rays.The procedure was resumed, without any delay, with a change in surgical technique, by using a ronjour.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
H11: h2: additional information on: d4 (lot number).
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H11: the associated device was returned and evaluated.The visual inspection revealed that the device is completely disassembled with the retainer ring not returned with the rest of the device.The entire device is worn from use.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.8MM GLENOID DRILL SHAFT
Common Device Name
BIT, DRILL
Manufacturer (Section D)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer (Section G)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key23900408
Report Number3002788818-2025-00141
Device Sequence Number5128516
Product Code HTW
UDI-Device Identifier00885556860236
UDI-Public885556860236
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDRS-0926-001
Device Lot NumberQJ2053
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2025
Supplement Date Received by Manufacturer01/06/2026
01/20/2026
Initial Report FDA Received Date12/26/2025
Supplement Report FDA Received Date01/08/2026
01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-