| Model Number |
LF-GP |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation and previous investigations, it likely suggests the alleged failure is due to probable causes such as damaged or deterioration of parts due to wear and tear, without any design or manufacturing issues.The most probable cause of this complaint is traced to expected or random component failure without any design or manufacturing issue.A root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the fiberscope exhibited dust inside lens causing a blurry image.There was no patient involvement.
|
| |
|