| Model Number |
VTICMO 12.6 (-14.5/ 3.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6- investigation type code: 4110- lens work order search-no similar complaint event(s) within associated lots were found.H11- manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced blurred vision and lens rotation/ repositioning.The lens was repositioned but that did not resolve the problem.In a separate surgery the lens was exchanged with a spherical model, same length and the problem was resolved.The cause of the event was reported as unknown.It has been noted that the surgeon selected a different lens size than that initially suggested by the icl calculation software.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: h3 - type of investigation code: 10- device evaluation: the lens was returned in a microcentrifuge tube with moisture and residue/debris on product.Visual inspection found haptic torn.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|