MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR; FOLDERS AND INJECTORS , INRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR; FOLDERS AND INJECTORS , INRAOCULAR LENS (IOL) Back to Search Results
Model Number IV
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-health care professional reported that during an intraocular lens (iol) implant procedure, the injector damaged the lens.Additional information was received and stated that, the lens was implanted and later explanted during initial procedure.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information was provided in h.3., h.6.And h.11.A sample was not received at the manufacturing site for evaluation for the report of a damaged lens; however, the attached customer photo confirms the reported issue of a damaged lens.Device/batch record review: a review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.The photo confirms the reported issue of a damaged lens; however, because an injector sample was not received at the manufacturing site, the root cause for the customer complaint issue cannot be determined.Based on the evaluation of the information received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.While the root cause and its origin are inconclusive, the following factors outlined in the reported product¿s instruction for use (ifu) could potentially contribute to an outcome similar to the reported event.The ifu states that the handpiece must be inspected prior to each use to confirm it is free of damage.It further emphasizes that if the handpiece¿particularly the plunger tip, appears damaged, bent, or otherwise unfit for proper use, it must not be used.In such cases, users are instructed to contact company immediately.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR
Common Device Name
FOLDERS AND INJECTORS , INRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23903605
Report Number2523835-2025-01347
Device Sequence Number7622091
Product Code MSS
UDI-Device Identifier00380659777745
UDI-Public(01)00380659777745(17)320630(11)220630(10)14V9L7
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K212039
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIV
Device Catalogue Number8065977774
Device Lot Number14V9L7
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/03/2025
Supplement Date Received by Manufacturer03/04/2026
Initial Report FDA Received Date12/28/2025
Supplement Report FDA Received Date03/30/2026
Date Device Manufactured06/30/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON VIVITY TORIC HYDROPHOBIC IOL.
Patient SexUnknown
-
-