• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN? EL ICD VR; Implantable Cardioverter Defibrillator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN? EL ICD VR; Implantable Cardioverter Defibrillator Back to Search Results
Model Number D151
Medical Device Problem Codes Premature Discharge of Battery (1057); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Moreover, the device emitted beeping tones.A request was made to have data from this device analyzed.Data analysis confirmed low voltage fault and that this device power consumption is increasing in a gradual fashion.A device replacement was recommended.Subsequently, this device was explanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN? EL ICD VR
Common Device Name
Implantable Cardioverter Defibrillator
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key23904177
Report Number2124215-2025-93303
Device Sequence Number21165605
Product Code LWS
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/24/2020
Device Model NumberD151
Device Catalogue NumberD151
Device Lot Number202057
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/02/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2025
Supplement Date Received by Manufacturer09/03/2026
Initial Report FDA Received Date12/29/2025
Supplement Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/05/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age66 YR
Patient SexMale
-
-