| Catalog Number |
03.501.080 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the devices are not pulling correctly; stiff.The lever that cuts is very wobbly.The issue was observed during pre-surgery and no injuries.There were no reports of injuries, medical intervention or prolonged hospitalization.It was reported that there was no further information regarding the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1.H3, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found normal traces of use were observed on the device.No nonconformances were observed.A dimensional inspection was not performed since it was not applicable to the complaint condition a functional evaluation was partially performed using the internal mechanisms of the applic-instr f/sternal zipfix and found device is able to be tightened without issues.As mating device was not returned, the reported event was unable to be replicated.The overall complaint was not confirmed as the observed condition of the applic-instr f/sternal zipfix would have not contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: part number: 03.501.080-us, lot number: 738p623, manufacturing site: hägendorf, release to warehouse date:27-april-2022.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9, h4.Corrected: g1.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H6: the product was not returned to depuy synthes; however, photos were provided for review.The photographs attached were reviewed, however they do not represent the reported complaint condition.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the photographs provided are not representative of the reported complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review.Part number: 03.501.080-us.Lot number: 738p623.Manufacturing site: hägendorf.Release to warehouse date:27-april-2022.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h6: the product was not returned to depuy synthes; however, photos were provided for review.The photographs attached were reviewed, however they do not represent the reported complaint condition.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the photographs provided are not representative of the reported complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review.Part number: 03.501.080-us.Lot number: 738p623.Manufacturing site: hägendorf.Release to warehouse date:27-april-2022.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|