• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ULTRA FAST-FIX AB ASSEMBLY - CURVED; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ULTRA FAST-FIX AB ASSEMBLY - CURVED; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Catalog Number 72201494
Medical Device Problem Code Firing Problem (4011)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that during a meniscus repair, when t1 was fired from one (1) ultra fast fix, both ts came out.The procedure was resumed, after a non significant delay, using an equivalent s+n back up.Additional anesthesia was not required as consequence of the surgical prolongation.No further complications were reported.
 
Additional Manufacturer Narrative
H11: h2: additional information on: b5 & h6 (impact code).
 
Event or Problem Description
It was reported that during a meniscus repair, when t1 was fired from one (1) ultra fast fix, both ts came out.T1 was deployed through the meniscus.Both ts were removed with tweezers.The patient's status is fine.The procedure was resumed, after a non significant delay, using an equivalent s+n back up.Additional anesthesia was not required as consequence of the surgical prolongation.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H11.H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found 3 ultra fast-fix ab assembly - curved.Only two of them appeared complete in the picture.One of them had the sutures with the implants attached outside the needle.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include an unintended second actuation of the device or excessive retraction of the device causing t2 to be pulled from the needle based on this investigation, the need for corrective action is not indicated.
 
Additional Manufacturer Narrative
H11: internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.A visual evaluation showed the device was returned in original packaging with the batch number of the complaint on the label.The t1, t2, and suture were returned on the distal end of the needle.The variable depth straw has been trimmed.An image evaluation was performed and found 3 ultra fast-fix ab assembly - curved.Only two of them appeared complete in the picture.One of them had the sutures with the implants attached outside the needle.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (an unintended second actuation of the device or excessive retraction of the device causing t2 to be pulled from the needle).Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA FAST-FIX AB ASSEMBLY - CURVED
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key23905933
Report Number1219602-2025-03329
Device Sequence Number13048418
Product Code GAT
UDI-Device Identifier03596010597229
UDI-Public3596010597229
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72201494
Device Lot Number2110687
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2025
Supplement Date Received by Manufacturer12/29/2025
02/09/2026
04/16/2026
Initial Report FDA Received Date12/29/2025
Supplement Report FDA Received Date12/30/2025
02/10/2026
04/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/20/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexMale
Patient Weight76 KG
-
-