MAUDE Adverse Event Report: HOLOGIC, INC 3DIMENSIONS; MAMMOGRAPHY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC 3DIMENSIONS; MAMMOGRAPHY SYSTEM Back to Search Results
Model Number 3DM-SYS-STD
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the user site that prior to a 3 dimensions patient exam on (b)(6) 2025, that the c-arm of the device moved down by itself.An error message was displayed on the device, in which the user site was unable to clear.No patient or user injuries were reported.A hologic field service engineer (fse) was dispatched to investigate the 3 dimensions device and observed that the device was functioning properly and additionally was unable to duplicate the adverse event.The fse reported that the error log of the device showed that right foot switch was stuck.The fse replaced both gantry foot switches, confirmed that all switches were working properly, and verified that the system is now operating according to manufacturer specifications.No further information is currently available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DIMENSIONS
Common Device Name
MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
600 technology drive
newark DE 19702
Manufacturer (Section G)
HOLOGIC, INC
600 technology drive
newark DE 19702
Manufacturer Contact
ugo chibueze
445 simarano drive
marlborough, MA 01752
MDR Report Key23907265
Report Number3005496266-2025-00122
Device Sequence Number10024165
Product Code OTE
UDI-Device Identifier15420045510524
UDI-Public(01)15420045510524(11)240312
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P080003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3DM-SYS-STD
Device Catalogue Number3DM-SYS-STD
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/29/2025
Supplement Report FDA Received Date06/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/12/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-