| Catalog Number |
IAB-05830-LWS-SC |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported " per the bpw assessment, iab appeared to be minimally inflating with significant kink noted.Most recent cxr ~10 hours prior.Recommended a stat cxr to confirm iab placement.Made the recommendation to hyperextend the patient's hip to see if there was any improvement to the bpw morphology.Bpw shows moderate kink with slow return to baseline.".Additional information states that the iabp console was swapped to continue therapy and the iab volume was decreased to continue therapy.No patient injury or consequence reported.The patient's current condition is reported as "fine".Please see associated mdr#: 3010532612-2025-01290.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The reported complaint for "helium loss alarms" is not able to be confirmed.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
|
| |
|
Event or Problem Description
|
|
It was reported " per the bpw assessment, iab appeared to be minimally inflating with significant kink noted.Most recent cxr ~10 hours prior.Recommended a stat cxr to confirm iab placement.Made the recommendation to hyperextend the patient's hip to see if there was any improvement to the bpw morphology.Bpw shows moderate kink with slow return to baseline.".Additional information states that the iabp console was swapped to continue therapy and the iab volume was decreased to continue therapy.No patient injury or consequence reported.The patient's current condition is reported as "fine".Please see associated mdr#: 3010532612-2025-01290, 3010532612-2025-01296.
|
| |
|
Search Alerts/Recalls
|