|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
The reported event is confirmed manufacturing related.This is addressed by capa 11092653.Visual evaluation noted received 1 all silicone foley catheter, attached to leaflet, syringe and drainage bag.Using solution attempted to inflate with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) but solution immediately leaked to the drainage funnel indicating lumen to lumen leak.Based on the physical sample received the reported event is confirmed manufacturing related and does not meet specifications per (b)(4) which states "catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe." the possible root causes for this failure, per capa 11092653, are "funnel wall thickness to thin due to molding core pin shape" or "extruded tube diameter with variation causing leak in catheter funnel molding." risk review, labeling review, and dhr review are not required as event has already been investigated per capa 11092653.Refer to capa 11092653 for further details.Correction: d, f, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|