| Catalog Number |
175814 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter name: (b)(6).The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that on insertion of urinary catheter, noticed leakage of urine coming out the actual catheter tube.A tear was found in the tube (unable to see before insertion as covered in the blue packaging upon insertion).Immediately removed the catheter and inserted a new catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states: warning: on catheter, do not use ointments or lubricants having a petroleum base.They will damage the catheter and may cause balloon to burst.This is a single use device.Do not resterilise any portion of this device.Reuse and or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and or lead to injury, illness or death of the patient.Directions for use visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Corrections: d, f, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that on insertion of urinary catheter, noticed leakage of urine coming out the actual catheter tube.A tear was found in the tube (unable to see before insertion as covered in the blue packaging upon insertion).Immediately removed the catheter and inserted a new catheter.
|
| |
|
Search Alerts/Recalls
|