| Catalog Number |
101/870/070CZ |
| Medical Device Problem Code |
Pressure Problem (3012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cuff pressure did not increase.The reporter noted that replacement was scheduled for one month on 10/21, but the cuff pressure did not increase on 11/7, so it was replaced.Then, on 11/28, the cuff pressure did not increase either, so it was replaced again.The event occurred during patient use/administration at patient's home.There was no reported patient harm/adverse event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for investigation.The reported issue was confirmed during functional testing when the reported issue was duplicated.The issue was traced to a hole found in the device cuff.As no lot number was provided, a review of device history records could not be conducted.A root cause could not be identified for the observed hole.
|
| |
|