• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. PACING SYSTEM ANALYZER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. PACING SYSTEM ANALYZER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 229047
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/29/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: it was determined at analysis that the analyzer was defective.It was noted that the analyzer displayed an error message stating that " analyzer backup battery was low".The analyzer was decommissioned as it was no longer serviceable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
The analyzer was returned for evaluation as an associate product and subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
 
Additional Manufacturer Narrative
Product analysis update: it was determined at analysis that the analyzer was non functional.It was noted that the analyzer displayed an error message stating that " analyzer backup battery was low".New battery had no effect and the error message remains.The analyzer was decommissioned as it was no longer serviceable.Correction: g3.Date mfr rec additional codes.Imf (annex f) health impact medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACING SYSTEM ANALYZER
Common Device Name
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key23919569
Report Number2182208-2025-13831
Device Sequence Number9041172
Product Code KRG
UDI-Device Identifier00721902256011
UDI-Public(01)00721902256011(11)020417
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number229047
Device Catalogue Number229047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2025
Supplement Date Received by Manufacturer12/16/2025
Initial Report FDA Received Date12/30/2025
Supplement Report FDA Received Date12/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/17/2002
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-