| Model Number |
CD3249-40 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Discomfort (2330)
|
| Date of Event |
12/10/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The cause of infection could not be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented with pain in the clinic with an infection at the implanted device pocket site.The physician explanted the entire system ¿ implantable cardioverter-defibrillator, right ventricular lead and left ventricle lead due to infection.The patient was in stable condition throughout the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned.Interrogation and visual inspections were normal.Analysis found that the device was received with battery voltage above elective replacement indicator (eri) level.The cause of infection could not be traced to the device.
|
| |
|