|
It was reported that an oily, waxy residue was found on multiple olympus enf-v3/rhino-laryngo videoscope and enf-vh/rhino-laryngo videoscope following reprocessing in the medivator automated endoscope reprocess.Primarily those reprocessed on (b)(6) 2025.The facility noted that alcohol removed the residue completely, while endozyme only partially improved it.No procedure delays were reported.There was no report of patient harm associated with this event.
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.The product was not returned to olympus for inspection; however, olympus technical assistance center (tac) provided remote troubleshooting for the oily residue that was found on scope.Troubleshooting was able to resolve the reported allegation.Olympus endoscopy support specialist (ess) was notified and engaged with the facility via email and phone, requesting additional information including affected model and serial numbers, point-of-use cleaning practices, detergent usage, and lubricant use.The facility confirmed that intercept detergent wipes were being used for point-of-use cleaning.Nursing staff were individually interviewed and confirmed to be following proper pre-cleaning procedures in accordance with the information for use (ifu).Additionally, an onsite visit was conducted by ess on december 4, 2025.The facility review summary confirmed that all reprocessing steps observed were performed in alignment with the olympus ifu across all categories, including procedure room setup, transportation and storage, reprocessing, care and handling, and facility layout.It was further reported that a significant improvement in the residue issue following the visit.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|