• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT ENDURITY MRI PACEMAKER, SR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT ENDURITY MRI PACEMAKER, SR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number PM1172
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the device is subject to the assurity and endurity pacemakers' header anomaly advisory issued by abbott on 15 march 2021.The pacemaker was explanted and replaced.The patient was stable.
 
Additional Manufacturer Narrative
The device was returned due to advisory with no allegation of a malfunction.Visual inspection of the device and header attachment area did not find any anomalies.Interrogation of the device indicated battery voltage and current within normal range when received.Based on this information, the device was found to communicate appropriately with a merlin programmer and has not reached the elective replacement indicator (eri).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY MRI PACEMAKER, SR
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key23921337
Report Number2017865-2025-1008300
Device Sequence Number9295740
Product Code LWP
UDI-Device Identifier05414734509596
UDI-Public(01)05414734509596(10)A000032593(17)190531
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
P140033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Model NumberPM1172
Device Lot NumberA000032593
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2025
Supplement Date Received by Manufacturer01/21/2026
Initial Report FDA Received Date12/30/2025
Supplement Report FDA Received Date01/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-