ARTHROCARE CORPORATION TRISTAR 50 ICW; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number ASC4630-01
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alter the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that during a surgery, the tristar 50 icw wand missed a piece of iron (with a diameter of about 0.4 centimeters).The procedure was completed; however, it is unknown if there was surgical delay.After the surgery, during instrument count, the missing piece was noted, and subsequent x-ray imaging did not reveal any fragments.Nevertheless, there remains a risk that the piece could have been retained inside the patient, despite the negative x-ray findings.
 
Event or Problem Description
It was reported that during a surgery, the tristar 50 icw wand missed a piece of iron (with a diameter of about 0.4 centimeters).The procedure was completed; however, it is unknown if there was surgical delay.After the surgery, x-ray imaging was performed and no fragment was found inside the patient.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H11 b5 and h6: event description, clinical codes and impact codes corrected.H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Factors that can contribute to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences.Based on this investigation, the need for corrective action is not indicated.If any additional information related to this event is received in the future, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRISTAR 50 ICW
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key23921835
Report Number3006524618-2025-00718
Device Sequence Number13085414
Product Code GEI
UDI-Device Identifier00817470001921
UDI-Public00817470001921
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K033584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberASC4630-01
Device Lot Number2173182
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2025
Supplement Date Received by Manufacturer01/15/2026
Initial Report FDA Received Date12/30/2025
Supplement Report FDA Received Date01/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
-
-