Catalog Number MCP496G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a thyroidectomy procedure on (b)(6) 2025 and suture was used.During the procedure, it was reported that the suture was broken when the surgeon attempted to sew the tissue.Changed another one to complete the surgery.There is no report on patient's injury.No additional information could be provided.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Corrected information: d 9.Device available for evaluation?, h 3.Device evaluated by manufacturer? additional information: d 9.Date device returned to manufacturer, d 9.Is device returned to manufacturer?, h 6.Component code, h 6.Type of investigation, h 6.Investigation findings, h 6.Investigation conclusions.Investigation summary the product was returned to ethicon for evaluation.One needle-suture piece outside its packaging was received for analysis.Product code mcp496g.During the initial inspection of the sample, no breakage suture was observed.Even though the extreme of the suture was noted to be cut likely by a surgical instrument.Body fluids were also observed on the suture.The other section of the suture was not returned for analysis.This product code mcp496g contains an absorbable suture.Since the sample was received open, the functional test could not be performed due to undetermined exposure time to the environment.Although no conclusion could be reached on the cause of the reported event.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of surgical instruments, such as needle holders and forceps.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCRYL+ UD 18IN 4-0 S/A PS-2 PRM MP
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key23921887
Report Number2210968-2025-14654
Device Sequence Number9082161
Product Code GAM
UDI-Device Identifier10705031046597
UDI-Public(01)10705031046597(17)270430(10)108H5S
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K050845
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMCP496G
Device Lot Number108H5S
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2025
Supplement Date Received by Manufacturer01/06/2026
Initial Report FDA Received Date12/30/2025
Supplement Report FDA Received Date01/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-