| Model Number |
DOC10 |
| Medical Device Problem Code |
Low Readings (2460)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant product: (b)(4): d-ys dy-s dura-y sensor x1, lot number- unknown.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the sensor connection cable and sensor produced low spo2 readings.The values ranged from 88 to 92%, although readings were expected to be above 95%.The sensor was used during a dental procedure in a separate room with no windows, and it was not covered.No error message appeared, only an alarm indicated low spo2 readings.The patient was an animal connected to an ecg and a blood pressure cuff and placed on a warming pad, but the sensor was also tested on a human, which also produced low readings.When another brand was used, results ranged from 97 to 100%.No patient harm occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the sensor connection cable and sensor produced low spo2 readings.The values ranged from 88 to 92%, although readings were expected to be above 95%.The sensor was used during a dental procedure in a separate room with no windows, and it was not covered.No error message appeared, only an alarm indicated low spo2 readings.The patient was an animal connected to an ecg and a blood pressure cuff and placed on a warming pad, but the sensor was also tested on a human, which also produced low readings.When another brand was used, results ranged from 97 to 100%.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|