DEPUY ORTHOPAEDICS INC US MBT CEM KEEL TIB TRAY SZ4; MBT TIBIAL TRAY : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 129433140
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint#: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: additional narrative: d6, h4: the information provided regarding the manufacturing date and implantation date are identified to be conflicting.Continued follow-up is being conducted and any additional information received will be reported.
 
Event or Problem Description
Thayer school of engineering at dartmouth first report of retrievals for statement of work for q4 of 2025.Dartmouth is the research location, and not the facility of placement.A depuy synthes implant was revised and reviewed for analysis.Reason for revision: revision for loosening of the femoral component, tibial component, and pain.The non-articular surface of the revised cemented tibial tray shows no cement remaining, and relative motion between bone and metal.The non-articular surface of the mobile-bearing polyethylene tibial insert shows severe abrasion.The articular surface demonstrates severe abrasion, pitting, and delamination.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT CEM KEEL TIB TRAY SZ4
Common Device Name
MBT TIBIAL TRAY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23923251
Report Number1818910-2025-22726
Device Sequence Number12989122
Product Code NJL
UDI-Device Identifier10603295025795
UDI-Public(01)10603295025795(17)251227(10)2664767
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/27/2025
Device Catalogue Number129433140
Device Lot Number2664767
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/11/2025
Supplement Date Received by Manufacturer01/01/2026
Initial Report FDA Received Date12/30/2025
Supplement Report FDA Received Date01/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-