• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ATTUNE CR FB INSRT SZ 7 5MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ATTUNE CR FB INSRT SZ 7 5MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 151620705
Medical Device Problem Codes Naturally Worn (2988); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Malfunction
Event or Problem Description
(b)(6) first report of retrievals for statement of work for q4 of 2025.(b)(6) is the research location, and not the facility of placement.A depuy synthes implant was revised and reviewed for analysis.Reason for revision: revision for loosening of the femoral component, tibial component, and pain.The articular surface shows minor yellowing and severe abrasion and pitting.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: b1 product problem, d10 concomitant, h1 type of reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FB INSRT SZ 7 5MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23923575
Report Number1818910-2025-22734
Device Sequence Number21085975
Product Code JWH
UDI-Device Identifier10603295045885
UDI-Public(01)10603295045885(17)240430(10)J37H08
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number151620705
Device Lot NumberJ37H08
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/11/2025
Supplement Date Received by Manufacturer02/09/2026
Initial Report FDA Received Date12/30/2025
Supplement Report FDA Received Date02/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/16/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 5 CEM; UNK ATTUNE FEMORAL
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-