Model Number 17026
Medical Device Problem Codes Break (1069); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual inspection confirms that the distal hexalobe tip has sheared off of the driver shaft.The root cause is likely the device reaching its fatigue limit during the application of final tightening.Review of device history records showed no manufacturing or processing related irregularities.Labeling review: warnings/cautions/precautions: risks identified with the use of these devices, which may require additional surgery, include device component failure, loss of fixation/stabilization, non-union, vertebral fracture, neurological injury, vascular or visceral injury.Possible adverse effects: initial or delayed loosening, disassembly, bending, dislocation, and/or breakage of device components.
 
Event or Problem Description
It was reported that the driver is damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVICTUS SPINAL FIXATION SYSTEM
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
ALPHATEC SPINE, INC.
1950 camino vida roble
carlsbad CA 92008
Manufacturer (Section G)
ALPHATEC SPINE, INC.
1950 camino vida roble
carlsbad CA 92008
Manufacturer Contact
michael hendrickson
1950 camino vida roble
carlsbad, CA 92008
7603566777
MDR Report Key23926189
Report Number2027467-2025-00286
Device Sequence Number9338557
Product Code HXX
UDI-Device Identifier00190376234175
UDI-Public(01)00190376234175(10)8525403
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number17026
Device Catalogue Number17026
Device Lot Number8525403
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2025
Initial Report FDA Received Date12/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-