SMITH & NEPHEW, INC. JOURN ART INS BCS STD 5-6 LT 9; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
74023261 |
| Medical Device Problem Codes |
Fracture (1260); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference case-(b)(4).H6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, after a tka surgery performed on (b)(6) 2013, the peg of the journ art ins bcs std 5-6 lt 9 fractured.A revision surgery was performed on (b)(6) 2025 to explant the fractured implant and replace it for s+n implants.The patient is still experiencing pain and discomfort after this revision surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, based on the limited information provided a clinical root cause for the broken peg on the poly cannot be confirmed.The first-generation journey bcs femoral and tibial insert components, were phased out globally in 2013-2014 as they may have a higher risk of requiring a revision earlier than expected.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed in warnings and precautions that the implant can break or become damaged as a result of strenuous activity or trauma.A review of the risk management file revealed this failure mode was previously identified.A review of previous actions concluded that there is a higher-than-expected risk of revision surgery for the first generation of journey bcs systems.It was recommended to undertake a field safety corrective action to inform surgeons of the higher expected risk of revision and ensure clear communication that the device should only be used for polyethylene exchange revision of journey bcs total knee constructs where the femoral component and tibial baseplate are well fixed.According with inspection drawing, the final inspection includes the verification of part configuration per print.Also, per material specification, the quality and manufacture of ultra-high-molecular-weight polyethylene (uhmwpe) as ram-extruded rod and compression-molded rod and plate shall be controlled.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include abnormal loading of limb, excessive forces and/or injury.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|