| Model Number |
420006-12 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The large needle driver instrument was analyzed and found to have a frayed grip cable at the idler pulleys.Some wires were broken, but the cable itself was not fully broken.The cable was spliced in one place.There was no discoloration, corrosion, or contamination observed on the cable that would occur from improper flushing or rinsing during reprocessing.Further inspection revealed no signs of missing parts and found no abnormally sharp or damaged components that could have contributed to the cable breaking.Common causes of frayed - distal instrument grip cable are attributed to damage to the cable through external collisions, during transport and/or storage, or during use or reprocessing.
|
| |
|