Block h6: imdrf device code a0504 captures the reportable event of a balloon leak in the esophagus.Block h11: investigation results.The returned cre pro wireguided dilatation balloon was analyzed, and a visual inspection found no damages to the device.Functional inspection was performed, and the balloon was able to be inflated without any problem; however, it was not able to hold the pressure due to it had a pinhole in the balloon which produces a leakage, located approximately at 83 mm from the tip of the device.Microscopic inspection confirmed the balloon pinhole.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon leak was confirmed.The pinhole found in the balloon which produces a leakage is likely to have occurred due to procedural factors such as excess of pressure, interaction with other devices, or anatomical conditions.Also, it is possible that interaction with a sharp surface during or previous to the procedure could have caused the problem found on the distal section of the balloon.Therefore, the most probable root cause is an adverse event related to procedure.A labeling review was performed, and from the information available, this device was used per the instructions for use (ifu)/product label.
|