BOSTON SCIENTIFIC CORPORATION CRE PRO; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00558650
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0504 captures the reportable event of a balloon leak in the esophagus.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre pro wireguided dilatation balloon was attempted to be used in the esophagus during a gastroscopy procedure to treat a stricture performed on (b)(6) 2025.During the procedure, the balloon was leaking.The procedure was completed with another cre pro wireguided dilatation balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0504 captures the reportable event of a balloon leak in the esophagus.Block h11: investigation results.The returned cre pro wireguided dilatation balloon was analyzed, and a visual inspection found no damages to the device.Functional inspection was performed, and the balloon was able to be inflated without any problem; however, it was not able to hold the pressure due to it had a pinhole in the balloon which produces a leakage, located approximately at 83 mm from the tip of the device.Microscopic inspection confirmed the balloon pinhole.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon leak was confirmed.The pinhole found in the balloon which produces a leakage is likely to have occurred due to procedural factors such as excess of pressure, interaction with other devices, or anatomical conditions.Also, it is possible that interaction with a sharp surface during or previous to the procedure could have caused the problem found on the distal section of the balloon.Therefore, the most probable root cause is an adverse event related to procedure.A labeling review was performed, and from the information available, this device was used per the instructions for use (ifu)/product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre pro wireguided dilatation balloon was attempted to be used in the esophagus during a gastroscopy procedure to treat a stricture performed on (b)(6) 2025.During the procedure, the balloon was leaking.The procedure was completed with another cre pro wireguided dilatation balloon.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE PRO
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key23931007
Report Number3005099803-2025-06948
Device Sequence Number18370080
Product Code FGE
UDI-Device Identifier08714729797531
UDI-Public(01)08714729797531(17)280110(10)35601313
Combination Product (Y/N)N
Initial Reporter CountryIC
PMA/510(K) Number
K112994
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00558650
Device Catalogue NumberM00558650
Device Lot Number0035601313
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2025
Supplement Date Received by Manufacturer02/13/2026
Initial Report FDA Received Date12/30/2025
Supplement Report FDA Received Date03/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
-
-