| Catalog Number |
8065751761 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
A customer reported that, during cataract surgery with intraocular lens (iol) implantation, an ophthalmic handpiece did not provide both torsional and longitudinal ultrasound.The surgery was completed on the same day without any patient harm.Additional information has been requested, none received till date.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in sections b5, h.6 and h.11.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.There was no product returned on this investigation.Based on the information obtained, the root cause of the reported event is inconclusive.A non-conformance-based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for ¿complaints reported against this serial number¿ was performed.There were no similar complaints reported for the product serial under investigation with the same event code.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Similar incident data was collected for the associated reported events.Quality assurance will continue to perform periodic monitoring for evidence of adverse trending and take further action as appropriate.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
Additional information indicates that the surgery was completed by replacing the ophthalmic handpiece.
|
| |
|