Model Number SN6AT5
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
A pharmacist reported that during an intraocular lens (iol) implant procedure, during insertion of lens into the eye, the injector pierced the lens.A backup lens was used to complete the procedure on the same day.Additional information has been requested.
 
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information was provided in d.9., h.3., h.6 and h.11.The lens was not returned in a delivery system as described.The lens was returned positioned incorrectly posterior side up in the lens case.Solution was dried on the lens.Both haptic tips were broken (not returned).No associated cartridge was returned for evaluation.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Balanced salt solution (bss) was indicated as the viscoelastic used.Which is non-qualified to be used in place of a viscoelastic inside a cartridge.The associated cartridge and handpiece were not provided.It is unknown if qualified products were used.The lens was not returned in a delivery system as described.The lens was returned positioned incorrectly posterior side up in the lens case.Solution was dried on the lens.Both haptic tips were broken.The root cause for the reported complaint is most likely related to a failure to follow the instructions for use (ifu).Bss was indicated as the viscoelastic used.Which is non-qualified to be used in place of a viscoelastic inside a cartridge.The associated cartridge and handpiece were not provided.It is unknown if qualified products were used.The ifu instructs: company foldable intraocular lens (iol) are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and or damage.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23932670
Report Number1119421-2025-03297
Device Sequence Number18544404
Product Code MJP
UDI-Device Identifier00380652251051
UDI-Public(01)00380652251051(17)300629(11)250701(10)16070459
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 02/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSN6AT5
Device Catalogue NumberSN6AT5.230
Device Lot Number16070459
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/11/2025
Supplement Date Received by Manufacturer01/30/2026
Initial Report FDA Received Date12/31/2025
Supplement Report FDA Received Date02/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BSS PLUS STERILE IRRIGATING SOLUTION; UNSPECIFIED INJECTOR
Patient Age77 YR
Patient SexUnknown
-
-