• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUREFIRE SCORPION NEEDLE; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUREFIRE SCORPION NEEDLE; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number SUREFIRE SCORPION NEEDLE
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 09-dec-2025, it was reported by an arthrex subsidiary employee via (b)(4) that an ar-13991n surefire scorpion needle tip was bent.It was reported that the scorpion failed to pass the suture, several attempts were made, then the scorpion needle was checked, and it was found that the tip where the notch is located was bent.This was discovered during use with no patient harm.Additional information has been requested.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.The complaint device was not received for evaluation.Based on the information provided, which may include the returned device (if available), pictures, videos, event descriptions, and any additional field data, arthrex concluded the most likely cause of the reported failure.The most likely cause of the reported failure is repetitive actuations, which can lead to bending of the needle tip.Additionally, the reported failure appears to be more consistent with an issue related to the scorpion suture passer.The complaint allegation was confirmed based on the evaluation of the customer-provided image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFIRE SCORPION NEEDLE
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23933108
Report Number1220246-2025-05901
Device Sequence Number15834576
Product Code MDM
UDI-Device Identifier00888867015920
UDI-Public(01)00888867015920(11)250512(17)300430(10)15434435
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUREFIRE SCORPION NEEDLE
Device Catalogue NumberAR-13991N
Device Lot Number15434435
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2025
Supplement Date Received by Manufacturer04/10/2026
Initial Report FDA Received Date12/31/2025
Supplement Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/12/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-