ZIMMER TMT TRABECULAR METAL FEMORAL METAPHYSEAL CONE, 35MM, MEDIUM, LEFT; PROSTHESIS, KNEE Back to Search Results
Catalog Number 00545002135
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); Aseptic loosening (4883)
Date of Event 09/08/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: medical product: trabecular metal femoral diaphyseal cone, 30mm, small, left: catalog#00545001730, lot#66108362; fem size d left: catalog#00588001401, lot#11025309; 11mm diameter 100mm length offset stem extension combined length 145mm: catalog#00598802011, lot#66379221; 12mm height size d articular surface with hinge post extension / use with plate 3,4,5,6 / screw enclosed do not discard: catalog#00588004012, lot#66815609; size 3 precoat cemented tibial component: catalog#00588000300, lot#66378437; trabecular metal tibial cone, large 60x36mm left: catalog#00545001360, lot#65531797; tibial central cone size fixed tibia: catalog#42545000508, lot#65761782.G2: foreign: australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation per hospital policy.The investigation is in process.Upon completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient underwent an initial total knee arthroplasty.Approximately nine (9) months post-implantation, the patient began to experience pain in the implanted joint.Subsequently, a revision surgery was performed due to loosening of the femoral components.All available information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, g3, g6, h2, h3, h6, h10, h11.Radiographic review identified mildly displaced periprosthetic lateral femoral condyle fracture and atypical alignment of the tibial trabecular metal cones, possibly caused by loosening.There was also mild valgus alignment of the left knee, a slightly short left femur and internal rotation of the tibia, which may have contributed to abnormal mechanics and predispose the patient for loosening.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRABECULAR METAL FEMORAL METAPHYSEAL CONE, 35MM, MEDIUM, LEFT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER TMT
10 pomeroy road
parsippany NJ 07054
Manufacturer (Section G)
ZIMMER TMT
10 pomeroy road
parsippany NJ 07054
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23933422
Report Number3005751028-2025-00025
Device Sequence Number15859900
Product Code MBH
UDI-Device Identifier00889024308466
UDI-Public(01)00889024308466(17)250531(10)64681621
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K103517
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2025
Device Catalogue Number00545002135
Device Lot Number64681621
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2025
Supplement Date Received by Manufacturer01/15/2026
Initial Report FDA Received Date12/31/2025
Supplement Report FDA Received Date01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age55 YR
Patient SexMale
-
-