| Model Number |
FASTPASS SCORPION, SL-MF |
| Medical Device Problem Codes |
Use of Device Problem (1670); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2025, it was reported by a sales representative via (b)(4) that 7 ar-13999mf scorpions jaws are not meeting together and cannot pass the suture.This was discovered during use in a case with a slight delay in the case.All procedures were completed successfully using free needles.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 one unpackaged ar-13999mf fastpass scorpion, sl-mf, batch 15455587, was received for investigation.Visual evaluation found that the jaw did not fully close as expected.Functional testing was not performed due to the device¿s condition.This is a known manufacturing issue, and an ncr will be opened to address it.The complaint allegation is confirmed.Refer to the investigation photographs.
|
| |
|