Catalog Number 1101300330
Medical Device Problem Codes Defective Component (2292); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 11/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient experienced skin burning and scabbing after xerf treatment on the face.A member of the cynosure lutronic clinical team reviewed the treatment parameters used and confirmed that they were within the normal approved clinical range.The patient reported that the burning and scabbing was experienced two days post treatment and lasted for seven days.It was reported that the patient has fully recovered.Review of the images provided of the device tip did observe a small burn mark on the effector surface.The tip has been returned to the manufacturer for further evaluation.At this time, it is unknown what contributed to the reported malfunction and burning observed damage to the tip observed in the provided image.It is unknown what caused this observed defect to the tip.The defect was observed but unable to be verified if it contributed to the reported burning.A final mdr will be submitted once investigation has been completed.Since a reported malfunction and potential for serious injury occurred, we are reporting this as an initial mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XERF 2.0
Common Device Name
XERF 2.0
Manufacturer (Section D)
LUTRONIC CORPORATION
lutronic center
219 sowon-ro
deogyang-gu, goyang-si 10534
KS  10534
MDR Report Key23933675
Report Number1222993-2025-00063
Device Sequence Number12168690
Product Code GEI
UDI-Device Identifier08809447657981
UDI-Public(01)08809447657981(11)251015
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/26/2026
Report Date (Section F) 06/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1101300330
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date12/01/2025
Event Location Outpatient Treatment Facility
Date Report to Manufacturer06/26/2026
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/31/2025
Supplement Report FDA Received Date06/26/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexFemale
Date Report Sent to FDA06/26/2026
-
-