• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3505700BC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 12/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a primary breast augmentation surgery with implantation of a 700cc mentor memorygel breast implant prosthesis.Post-operatively, the patient was diagnosed with a ruptured right implant using mammogram imaging.As a result, the patient underwent implant removal on (b)(6) 2025.
 
Additional Manufacturer Narrative
On january 07, 2026, the mentor analysis lab received the suspect medical device for evaluation.On january 14, 2026, mentor completed an evaluation on the returned device.Device evaluation: mentor conducted a visual inspection.Visual analysis of the returned device revealed that the breast implant was found to be ruptured.As the authorization form for examination was not received, the product evaluation lab could not reach a conclusion regarding the specific nature of the ruptures.The evaluation was limited to non-destructive testing.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On january 28, 2026, mentor was notified that the patient was noted to have some breast asymmetry in addition to the rupture.The patient underwent bilateral exchange with mentor gel implants during the removal surgery.On january 30, 2026, mentor completed a reevaluation on the returned device as the authorization form for examination was received.Addition to device evaluation: mentor conducted a visual inspection, microscopic examination, and thickness measurement of the device.Visual analysis of the returned device revealed that the breast implant was found to be ruptured and received in three (3) parts.Microscopic examination was performed, and the cause of the tear could not be identified.Therefore, a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.No conclusion could be reached on the cause of the reported event.Asymmetry may be attributed to one or more of the following: contracture of the fibrous capsule, seroma or hematoma, development of postoperative breast dysplasia, unilateral discrepancy in muscle development, deflation of the implant, incorrect choice of implant shape or size, and surgical technique.Asymmetry is a known complication associated with these devices and is referenced in our current product insert data sheet.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key23933894
Report Number1645337-2025-14467
Device Sequence Number9627341
Product Code FTR
UDI-Device Identifier00081317007012
UDI-Public(01)00081317007012(17)221203(10)7528288
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/03/2022
Device Catalogue Number3505700BC
Device Lot Number7528288
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/05/2025
Supplement Date Received by Manufacturer01/07/2026
01/28/2026
Initial Report FDA Received Date12/31/2025
Supplement Report FDA Received Date01/23/2026
02/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
Patient SexFemale
-
-