• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PROXIMATE*PLUS MD SKIN STPLR; REMOVABLE (SKIN)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PROXIMATE*PLUS MD SKIN STPLR; REMOVABLE (SKIN) Back to Search Results
Catalog Number PMW35
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the nephrectomy surgery, after fired, noted the staples malformed.Changed the new one to complete surgery, the device was damaged.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 12/31/2025.D4: batch#: unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 1/21/2026.Investigation summary: the product was returned to ees for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that the pmw35 device was returned nonfunctional.The device was received with the cartridge partially detached from the handle due to insufficient handle weld on the upper handle directors.No functional test was performed due to the condition of the device; however, the insufficient handle weld could have cause firing issues when the cartridge began to be detached from the handle.The separation of the device components has been correlated to the manufacturing process, and a product issue was identified during the investigation of the sample received.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ees quality system.Device history review: a manufacturing record evaluation was performed for the finished device lot 238d81, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
REMOVABLE (SKIN)
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
3035526892
MDR Report Key23934386
Report Number3005075853-2025-09879
Device Sequence Number17549211
Product Code GDT
UDI-Device Identifier10705036002642
UDI-Public(01)10705036002642(17)290731(10)238D81
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K833357
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPMW35
Device Lot Number238D81
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/08/2025
Supplement Date Received by Manufacturer01/05/2026
Initial Report FDA Received Date12/31/2025
Supplement Report FDA Received Date01/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-