• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC FLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATORS; FED ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC FLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATORS; FED ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number G34874
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1 - phone number: (b)(6).H3 - device evaluation anticipated but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported that the sheath liner from a flexor parallel ureteral access sheath and dilators delaminated during an unknown procedure.When inserting the flexible ureteroscope into the ureteral access sheath, force was used due to resistance being felt and the inner liner flaked off resulting in the fibers detaching in the patient.The foreign fibers were removed using an additional device.The patient is being closely monitored.Additional information regarding the event and patient outcome has been requested but is currently unavailable.
 
Additional Manufacturer Narrative
Additional information: d4 (lot #, expiration date, primary udi number), h4.H3 - device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATORS
Common Device Name
FED ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key23936251
Report Number1820334-2025-01616
Device Sequence Number9611314
Product Code FED
UDI-Device Identifier00827002348745
UDI-Public(01)00827002348745(17)280424(10)16573458
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K172217
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG34874
Device Catalogue NumberFUS-107045-P
Device Lot Number16573458
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2025
Supplement Date Received by Manufacturer03/12/2026
06/19/2026
Initial Report FDA Received Date12/31/2025
Supplement Report FDA Received Date03/23/2026
06/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DORMIA INSTRUMENT.
Patient SexUnknown
-
-