| Model Number |
9500-160 |
| Medical Device Problem Code |
Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).E1 initial reporter state (b)(6).
|
| |
|
Event or Problem Description
|
|
Dexcom became aware that the user experienced a failure but did not receive the expected sensor failure alert on their g7 cgm application.Data was provided and an investigation was performed.The root cause was determined to be a coding issue.The g7 cgm application only detects that the session has ended, thus, it does not alert the user that the sensor has failed.No injury or medical intervention was reported.
|
| |
|
Search Alerts/Recalls
|