| Model Number |
TPL0059 |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information regarding a guidance system being used outside of a procedure.It was reported that defect-23633 was identified.For the user to reproduce the issue they must perform all necessary steps and reach operation screen, open the review planning screen, select an implant (for example - right screw), and while the implant was selected, switch the control to surgeon screen.In the surgeon screen the user must press on the same implant, which results in the "tools" tab disappearing when switching control between the workstation and the surgeon screen.When pressing the selected implant - the tools tab does not open.The user needs to select a different implant for the "tools" tab to reappear.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H7) this regulatory report is being submitted due to retrospective review through fa1523, 806 report # 3005075696-12/16/2025-001-c.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: product id: unk_mxse_sw, software version: unk.H3, h6) analysis was performed for this event.In summary, a software defect was confirmed.Codes b01, c10, and d18 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: information regarding the software version was provided.Software has been updated to the below.Product id: kit1378-06 software version: 5.1.2 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|