• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIRLIFE VITAL SIGNS; EXTENSION SET FOR ADULT ANESTHESIA BREATHING CIRCUIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIRLIFE VITAL SIGNS; EXTENSION SET FOR ADULT ANESTHESIA BREATHING CIRCUIT Back to Search Results
Model Number 5108EX
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code Low Oxygen Saturation (2477)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product involved in the report has not been returned for evaluation.A review of the device history record is not possible as no lot number was provided.The documented udi is based on the stock/product code provided by the reporter; the udi-pi is not available as no lot number was provided all information reasonably known as of 01 jan 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as complaint-(b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
 
Event or Problem Description
Sunmed holdings llc.Received a single report that referenced three different incidences, which were associated with separate units, involving three different patients.This is the first of three reports.Refer to 1314417-2026-00002 for the second report.Refer to 2028807-2026-00001 for the third report.It was reported, the circuit disconnected multiple times, which has resulted in the patient desaturating, down into the 60s.No information was provided concerning medical intervention or patient outcome.
 
Additional Manufacturer Narrative
Additional information-investigation completion: h6.The product involved in the report was not returned for evaluation; however, an airlife team was onsite to discuss the complaint with the hospital staff.During the meeting, the clinical staff was unable to identify a specific disconnection point when questioned; therefore, several different connection points were reviewed; during the review, a connection point between the blue cuff and the circuit tubing was observed to be below specification.It was noted that this is the most likely location for the disconnection; therefore, the reported event of elbow//disconnection is considered to be confirmed.Manufacturing processes were reviewed and updated, to replace the elbow with the observed defect with an elbow with an alternative material and a tethered luer cap.A review of the device history record is not possible as no lot number was provided.The documented udi is based on the stock/product code provided by the reporter; the udi-pi is not available as no lot number was provided all information reasonably known as of 06 feb 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
 
Event or Problem Description
Sunmed holdings llc.Received a single report that referenced three different incidences, which were associated with separate units, involving three different patients.This is the first of three reports.Refer to 1314417-2026-00002 for the second report.Refer to 2028807-2026-00001 for the third report.It was reported, the circuit disconnected multiple times, which has resulted in the patient desaturating, down into the 60s.No information was provided concerning medical intervention or patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITAL SIGNS
Common Device Name
EXTENSION SET FOR ADULT ANESTHESIA BREATHING CIRCUIT
Manufacturer (Section D)
AIRLIFE
2710 north ridge dr nw
ste a
grand rapids MI
Manufacturer (Section G)
WEST MED
carretera internacional
salida norte no. 1053
magdalena, mx 84160
MX   84160
Manufacturer Contact
racheal cooper
2710 north ridge dr nw
ste a
grand rapids, MI 
MDR Report Key23940762
Report Number1314417-2026-00001
Device Sequence Number9215503
Product Code CAI
UDI-Device Identifier10190752118775
UDI-Public(01)10190752118775
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5108EX
Device Catalogue Number5108EX
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2025
Supplement Date Received by Manufacturer01/12/2026
Initial Report FDA Received Date01/01/2026
Supplement Report FDA Received Date02/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-