| Model Number |
CCWET3 |
| Medical Device Problem Code |
Use of Incorrect Control/Treatment Settings (1126)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an intraocular lens (iol) implant procedure, lens was explanted at secondary procedure due to patient experienced blurred vision difficulty focusing with dominant eye the clinical reason for explant was reported to be miscalculation of toric correction due to pre op measurements.The iol was replaced with unspecified lens approximately three months after initial procedure.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|