• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. NEXIVA DIFFUSICS; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. NEXIVA DIFFUSICS; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383596
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd nexiva diffusics w/maxzero 22g x 1.00 in leaked it was reported by customer that leaking at connection of maxzero to hub of diffusics.Small leakage/movement of fluids out of the threads and onto patient or hcp.Mainly on flushing.Leaking at connection of maxzero to hub of diffusics.Small leakage/movement of fluids out of the threads and onto patient or hcp.Mainly on flushing.
 
Additional Manufacturer Narrative
Investigation results: our quality engineer inspected the 2 photos and 2 used physical samples submitted for evaluation.The reported issue of leakage was not confirmed upon inspection of the samples.Analysis of the sample showed that visual inspection was performed and no damage was observed.A leak test was also performed on the 2 physical samples and no leakage was observed.Bd could not determine a manufacturing related root cause since the reported defect was not confirmed during the evaluation of the sample.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Production records were reviewed, and this batch meets our manufacturing specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXIVA DIFFUSICS
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key23944228
Report Number9610847-2025-00483
Device Sequence Number12723998
Product Code FOZ
UDI-Device Identifier00382903835966
UDI-Public(01)00382903835966(11)251103(17)280930(10)5281696
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K233529
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number383596
Device Lot Number5281696
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2025
Supplement Date Received by Manufacturer02/03/2026
Initial Report FDA Received Date01/02/2026
Supplement Report FDA Received Date02/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-