• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR INC. HI-TORQUE COMMAND¿; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR INC. HI-TORQUE COMMAND¿; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 2078175
Medical Device Problem Codes Break (1069); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a stenosed lesion in the common femoral artery (cfa) with calcification.Three ht command es 300cm guide wires (gw) were used in the procedure and resistance was noted during advancement of the guide wires with the anatomy.The guide wires were difficult to track and became kinked.Additionally, the gw tips broke but remained in one piece.A 4th ht command gw was able to continue the procedure without issue.There were no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.The additional devices referenced in b5 are filed under separate medwatch report numbers.
 
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot of specific issue.Based on the information provided, a definitive cause for the reported issue could not be determined.The difficulty to advance may be related to anatomical conditions.In certain circumstances, such as manipulation through tortuous and/or tight lesions, where heavy torquing and/or pushing/pulling is required, damage to a wire including a break may occur.Additionally, tortuosity, or over torquing may have damaged the device during advancement.Wire damage will lead to difficulty to advance.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE COMMAND¿
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 2807
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005737652
road no.2 km 58.0 cruce dávila
barceloneta PR 00617
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key23944491
Report Number2024168-2026-00017
Device Sequence Number12724127
Product Code DQX
UDI-Device Identifier08717648176685
UDI-Public(01)08717648176685(11)250915(17)270831(10)5091572
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K122573
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2078175
Device Catalogue Number2078175
Device Lot Number5091572
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2025
Supplement Date Received by Manufacturer01/02/2026
Initial Report FDA Received Date01/02/2026
Supplement Report FDA Received Date01/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-