MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL HD EPSCP,4.0,30,167,MITEK; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL HD EPSCP,4.0,30,167,MITEK; ARTHROSCOPE Back to Search Results
Catalog Number 242018
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a rotator cuff repair procedure the hd epscp, 4.0, 30, 167, mitek device produced a very blurry image.It was not reported if there were any delays to the surgical procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: investigation summary the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: repair level 3 major repair - other damages caused by customer - outer tube damaged, distal tip distal tip damaged dings/ scratched - illumination distal tip fiber damaged - particulate, optics particulate under distal lens particulate in system - optical system, optical components distal cover glass/negative damaged scratched/residue - cosmetic issue scratches/dents visual : scratched/nicked, broken (2+ pieces) : chipped/shaved/delaminated, visual : deformed/bent and labeling : illegible etch per service reports, this complaint was confirmed.Based on the investigation findings, the potential root cause is traced to faulty parts.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD EPSCP,4.0,30,167,MITEK
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle MA CH-24 00
SZ   CH-2400
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key23945496
Report Number1221934-2026-00028
Device Sequence Number10757730
Product Code HRX
UDI-Device Identifier10886705029150
UDI-Public(01)10886705029150
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K080560
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number242018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2025
Supplement Date Received by Manufacturer02/04/2026
Initial Report FDA Received Date01/02/2026
Supplement Report FDA Received Date02/04/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-