| Catalog Number |
04P53-25 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer observed false repeat reactive architect hbsag qualitative results which questioned by the physician for several patients.The one result example provided were: initial=1.23 s/co (> or =1.00 s/co=reactive) /repeated after redrawn multiple times same patient=1.6, 1.7 and 1.33 s/co hbsag qual confirmatory c2=1.55 s/co; % neutralization=99% (c2=> or =0.70 s/co and % neutralization=> or =50%=confirmed positive) additional information provided: all other hepatitis b markers are negative (no actual results provided) there was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint evaluation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history record review for architect hbsag qualitative reagent, lot 78397fz00.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.Review of tracking and trending for the architect hbsag qualitative assay did not identify an increase in complaint activity related to the complaint issue.Additionally, an increase in complaint activity was not identified for reagent lot 78397fz00.Device history record review did not identify any non-conformances or deviations with the complaint lots and complaint issue.A review of the product labeling concluded that the issue is sufficiently addressed.There was no issue with reagent performance identified.Based on the investigation, the device met performance specification and performed as intended at the customer site, therefore there is no systemic issue or deficiency with the architect hbsag qualitative reagent, lot 78397fz00.
|
| |
|
Event or Problem Description
|
|
The customer observed false repeat reactive architect hbsag qualitative results which questioned by the physician for several patients.The one result example provided were: initial=1.23 s/co (> or =1.00 s/co=reactive) /repeated after redrawn multiple times same patient=1.6, 1.7 and 1.33 s/co.Hbsag qual confirmatory c2=1.55 s/co; % neutralization=99% (c2=> or =0.70 s/co and % neutralization=> or =50%=confirmed positive).Additional information provided: all other hepatitis b markers are negative (no actual results provided).There was no reported impact to patient management.
|
| |
|