• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number TE4043A
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 12/09/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H.6.Investigation conclusions code d15: there is insufficient information available for gore to reasonably draw conclusions related to aspects of the event, therefore conclusion code ¿d15: cause not established¿ is being used.Insufficient information may include limited or missing relevant medical records, involvement of multiple implanted devices (including non-gore devices) in the field of treatment, patient non-compliance, and/or a general lack of available detail or specificity related to an adverse event and/or device.H.6.Type of investigation code b20: the device remains implanted and is not available for analysis.H.6.Investigation findings code c19: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2025, a patient was treated for a dissection using a gore® tag® thoracic branch endoprosthesis.On (b)(6) 2025, the patient was treated for a dissection using a gore® excluder® thoracoabdominal branch endoprosthesis (tambe) system, including gore® viabahn® vbx balloon expandable endoprosthesis devices.On follow-up ct, the physician noted a proximal endoleak on the tbe device.On (b)(6) 2025, a reintervention occurred.A tbe extender was placed proximally of the tbe device.After deployment, the physician attempted to withdraw the catheter, but it felt like it was stuck.The physician pushed the catheter forward, and when this occurred, the deployed device also pushed forward, covering 80% of the innominate artery.The physician performed a left subclavian - left carotid bypass, and then placed a vbx stent in the innominate artery.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
Updated h.6.Investigation conclusions from d12 to d1101.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
bryce wargin
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key23946006
Report Number2017233-2026-07007
Device Sequence Number10683229
Product Code MIH
UDI-Device Identifier00733132654338
UDI-Public(01)00733132654338(17)271207
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P210032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue NumberTE4043A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/02/2026
Supplement Report FDA Received Date03/05/2026
Date Device Manufactured12/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
Patient SexMale
-
-