The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.A remstar pro c-flex+ device was returned to a third-party service center for evaluation.There was no report of serious or permanent harm or injury.There was no report of medical intervention.During the evaluation of the device, the third-party service center observed evidence of foam particles inside the blower kit.In addition, the device had dust/dirt contamination, fluid contamination, and displayed error code/ displayed error code e050 (err_daily_values_corrupt),error code e053 (err_comp_log_sem_timeout), error code e055 (err_therapy_queue_full), and error code e063 (err_stuck_knob_key).The device was scrapped as per customer request.
|