| Model Number |
6662 |
| Medical Device Problem Code |
Low impedance (2285)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
B3: event date is estimated.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's system is exhibiting low impedances.Surgical intervention may be pending to address this issue.
|
| |
|
Event or Problem Description
|
|
Additional information received reports that the patient underwent surgical intervention on (b)(6) 2026, in which the ipg was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported observation against the returned ipg of low impedances was not confirmed.The root cause of the reported observation was not ascertained; the device exhibited normal device characteristics when using clinicians programmer and performing a system impedance test all ipg electrodes measured within the expected range.The ipg was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.The ate test specifically tests the ipg output and system impedances on all 16 electrodes including the ipg can connection.
|
| |
|
Search Alerts/Recalls
|