| Model Number |
VTICM5_12.1 (-10.50/2.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/17/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Initial mdr was submitted in error and is not applicable.Claim: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Secondary surgical intervention to remove/replace/reposition is identified in the labeling as a known potential adverse event(s) following icl implantation.H6: health effect clinical code 4581: enlargement of incision.H6: investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Claim: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that during an implantable collamer lens (icl) implantation procedure, the patient experienced poor fixation which caused incision expansion.The lens was implanted & removed intraoperatively.A replacement lens implanted at a later date & problem resolved.Cause of the event listed as patient related factor.
|
| |
|
Search Alerts/Recalls
|